The R&S FSUP26 Signal Source Analyzer combines the functionality of a high performance spectrum and signal analyzer with the advantages of a phase noise test system, all in a single box. The R&S FSUP26 is a unique and simple-to-use one-box solution for measuring oscillators and synthesizers in development and production applications. In addition, the Rohde & Schwarz FSUP26 leads to enormous cost reductions.
The Rohde and Schwarz FSUP26 analyzer is perfect for not only modern communications and broadcast systems, but also radar applications. Use the FSUP26 to measure phase noise and when characterizing oscillators, measure tuning slope, transient response, power, harmonics and spurious emissions, as well as amplifier noise. The integrated DC sources enable a wide range of additional measurements.
Features and specifications of:
Frequency range up to 26.5 GHz (up to 110 GHz with external mixers)
Low Noise DC outputs for supply and tuning voltages
Maximum flexibility for phase noise measurements
Oscillator characterization
Analysis of digital and analog modulated signals
Capabilities of the FSUP26 include:
Phase detector method with internal/external reference
Two-DUT method
High sensitivity
Automatic setting of all important parameters
Easy operation
Detection, suppression and listing of interference
Measurement of reference points as a function of frequency
Measurement of residual phase noise
AM noise
Abbreviated specifications of the FSUP26:
Operating Modes:
Signal Source Analyzer: 1 MHz – 26.5 GHz
Spectrum Analyzer: 20 Hz – 26.5 GHz
Signal Source Analyzer
Phase Noise Measurement:
W/ Spectrum Analyzer: 10 MHz – 26.5 GHz
W/ Phase Comparator: 1 MHz – 26.5 GHz
Internal reference: 1 MHz – 26.5 GHz
External reference: 1 MHz – 8 GHz
W/ Phase Comparator & Cross-Correlation: 1 MHz – 26 GHz
Transient Measurement:
Min. frequency offset: 10 MHz
Max. frequency offset: 30 MHz
Residual Noise w/ Phase Comparator: 1 MHz – 8 GHz
AM Noise: 1 MHz – 8 GHz
Sensitivity with internal reference and internal phase detector:
Input level >+10 dBm, spurious and harmonics <–30 dBc, mode “averaged”, +20 °C to +30 °C. LNA gain 40 dB, loop bandwidth =10 × frequency offset, max. 1 kHz. With the R&S Option B60 low phase noise Option B61 correlation extension option.
|
Freq Offset |
Input frequency, values in dBc (1Hz) |
|
5 MHz |
10 MHz |
100 MHz |
1 GHz |
3 GHz |
7 GHz |
10 GHz |
20 GHz |
|
|
1 Hz |
-114 |
-116 |
-87 |
-75 |
-62 |
-55 |
-52 |
-48 |
|
10 Hz |
-136 |
-135 |
-110 |
-91 |
-87 |
-80 |
-77 |
-71 |
|
100 Hz |
-143 |
-146 |
-134 |
-115 |
-106 |
-97 |
-95 |
-89 |
|
1 kHz |
-157 |
-161 |
-160 |
-134 |
-123 |
-118 |
-116 |
-112 |
|
10 kHz |
-165 |
-168 |
-168 |
-143 |
-131 |
-129 |
-126 |
-120 |
|
100 kHz |
-171 |
-170 |
-176 |
-158 |
-139 |
-140 |
-138 |
-132 |
|
1 MHz |
– |
-175 |
-177 |
-165 |
-160 |
-155 |
-150 |
-146 |
|
10 MHz |
– |
– |
-179 |
-172 |
-170 |
-170 |
-167 |
-161 |
|
30 MHz |
– |
– |
-179 |
-172 |
-170 |
-170 |
-170 |
-165 |
VCO Parameter Characterization:
| Measurement Parameters | VCO tuning characteristic, VCO tuning slope, power, pushing ON/OFF, harmonics measurement, VCO DC characteristic, summary | |
|
Frequency Range |
10 MHz – 26.5 GHz |
|
|
Power Supplies |
tuning ports |
2 |
|
DC ports |
2 |
|
|
additional ports |