The Keysight DSOX3032A (Agilent) is a two-channel model in the InfiniiVision 3000 X-Series. Engineered with breakthrough technology this upgradable oscilloscope provides superior performance and functionality. Its unique 5-instruments-in-one design delivers more scope for the same budget.
Features and Specifications of:
Features of:
●One million waveforms per second update rate
●MegaZoom IV smart memory technology
●Large 8.5-inch WVGA display
●Optional segmented memory
●Industry’s first 5-instruments-in-1:
oOscilloscope
oDigital channels
oBuilt-in 20 MHz function/arbitrary waveform generator with modulation
oIntegrated digital voltmeter
oProtocol analyzer
●Investment protection with Industry’s only fully-upgradable oscilloscope, including bandwidth up to 1 GHz
●Industry’s leading applications solution
Specifications of the Keysight DSOX3032A include:
●Bandwidth (-3 dB): 350 MHz
●Calculated rise time (10 – 90%): ≤ 1 ns
●Input channels: 2 analog channels
●Maximum sample rate:
o4 GSa/s half channel
o2 GSa/s all channel
●Maximum memory depth:
oStandard 2 Mpts
oOptional 4 Mpts
oOptional segment memory
●Display size and type: 8.5 WVGA display
●Waveform update rate: > 1 million waveforms per second
●Number of active probes supported: In general, one for 2-channel models
●Vertical system analog channels:
oHardware bandwidth limits: Approximately 20 MHz (selectable)
oInput coupling: AC, DC
oInput impedance: Selectable:
■1 MΩ ± 1% (14 pF)
■50 Ω ± 1.5%
oInput sensitivity range: 1mV/div to 5 V/div (1 MΩ and 50 Ω)
oVertical resolution: 8 bits (measurement resolution is 12 bits with averaging)
oMaximum input voltage:
■300 Vrms, 400 Vpk
■Transient overvoltage 1.6 kVpk
■With N2862A, N2863A, N2890A 10:1 probe: 300 Vrms
■Frequency de-rating (assumes sine wave input): 400 Vpk until 40 kHz, then de-rates at 20 dB/dec until 6 Vpk
oDC vertical accuracy: ±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
oDC vertical gain accuracy: ± 2.0% full scale
oDC vertical offset accuracy: ± 0.1 div ± 2 mV ± 1% of offset setting
oChannel-to-channel isolation: > 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and coupling on channels)
oOffset range:
■± 2 V (1 mV/div to 200 mV/div)
■± 50 V (> 200 mV/div to 5 mV/div)
●Vertical system digital channels:
oDigital input channels: 16 digital (D0 to D15. pod 1:D7 ~ D0, Pod 2: D15 ~ D8)
oThresholds: Threshold per pod
oThreshold selections: TTL (+1.4 V), 5V CMOS (+2.5 V), ECL (-1.3 V), user-defined (selectable by pod)
oUser-defined threshold range: ± 8.0 V in 10 mV steps
oMaximum input voltage: ± 40 V peak CAT l; transient overvoltage 800 Vpk
oThreshold accuracy: ± (100 mV + 3% of threshold setting)
oMaximum input dynamic range: ± 10 V about threshold
oMinimum voltage swing: 5 mVpp
oInput impedance: 100 kΩ ± 2% at probe tip
oInput capacitance: ~8 pF
oVertical resolution: 1 bit
●Horizontal system analog channels:
oTime base range: 2 ns/div to 50 s/div
oTime base accuracy: 25 ppm ± 5 ppm per year (aging)
oTime base delay time range:
■Pre-trigger: Greater of 1 screen width or 250 μs
■Post-trigger: 1 s to 500 s
oChannel-to-channel deskew range: ± 100 ns
oΔ Time accuracy (using cursors): ± (time base acc. x reading) ± (0.0016 x screen width) ± 100 ps
oModes: Main, zoom, roll, XY
oXY:
■On channels 1 and 2 only, Z Blanking on ext trigger input, 1.4 V threshold
■Bandwidth: Maximum bandwidth
■Phase error at 1 MHz: < 0.5 degree
●Horizontal system digital channels:
oMinimum detectable pulse width: 5 ns
oChannel-to-channel skew:
■2 ns (typical)
■3 ns (maximum)
●Aquisition system:
oMaximum analog channels sample rate:
■4 GSa/s half channel interleaved
■2 GSa/s all channel
oMaximum analog channels record length:
■2 Mpts half channel interleaved, 1 Mpts all channel (standard)
■4 Mpts half channel interleaved, 2 Mpts all channel (optional with DSOX3MEMUP (-040))
oMaximum duration of time captured at highest sampling rate (all analog channels): 500 us with 4M memory upgrade
oMaximum digital channels sample rate: 1 GSa/s
oMaximum digital channels record length:
■1 Mpts (standard with digital channels only)
■2 Mpts (optional with DSOX3MEMUP – with digital channels only)
oModes:
■Normal: Default mode
■Peak detect: Capture glitches as narrow as 250 ps at all time base settings
■Averaging: Selectable from 2, 4, 8, 16, 64, … to 65,536
■High resolution:
■Real time boxcar averaging reduces random noise and effectively increases vertical resolution
■12 bits of resolution when ≥ 10 μs/div at 4 GSa/s
■or ≥ 20-μs/div at 2 GSa/s
■Segmented:
■Segmented memory optimizes available memory for data streams that have long dead times between activity
■Maximum segments = 1000
■Re-arm time = 1 μs (minimum time between trigger events)
●Trigger system:
oTrigger sources: Analog channel (1~4), digital channel (D0~D15), line, external, WaveGen (1 or mod) (FM/FSK)
oTrigger modes:
■Normal (triggered): requires trigger event for scope to trigger
■Auto : triggers automatically in absence of trigger event
■Single: triggers only once on a trigger event, press [Single] again for scope to find another trigger event, or press [Run] to trigger continuously in either Auto or Normal mode
■Force: front panel button that forces a trigger
oTrigger coupling:
■DC: DC coupled trigger
■AC: AC coupled trigger, cutoff frequency: < 10 Hz (internal);
■High frequency reject: cutoff frequency ~ 50 kHz
■Low frequency reject: cutoff frequncy ~ 50 kHz
■Noise reject:
■Selectable OFF or ON
■Decreases sensitivity 2x
oTrigger holdoff range: 40 ns to 10.00 s
oTrigger sensitivity:
■Internal: < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
■External:
■200 mVpp from DC to 100 MHz
■350 mVpp 100 MHz to 200 MHz
oTrigger level range:
■Any channel: ± 6 div from center screen
■External: ± 8 V